Specifications

B-Scan Modes

Four sets of electronic distance measurement calipers with variable velocity

  • Two sets of electronic angle measurement calipers (variable velocity)
  • Movie sequence - real-time viewing and editing capability

10 MHz Posterior Segment

40 MHz UBM Wide-Field Anterior Segment

  • 25 frames-per-second image acquisition rate
  • 13 frames-per-second image acquisition rate
  • 10-second movie loop capability
  • 20-second movie loop capability
  • Sealed probe
  • Adjustable transmit gain (minimum to 0 dB)
  • Adjustable transmit gain (minimum to 0 dB)
  • Adjustable receive gain (27-90 dB)

  • Adjustable receive gain (27-90 dB)
  • Adjustable dynamic range (Log, S1, S2, S3)
  • Adjustable dynamic range (Log, S1, S2, S3)

  • Axial resolution: 23 microns
  • Axial resolution: 50 microns
  • Lateral resolution: 33 microns
  • Lateral resolution: 100 microns
  • Scanning angle: 30 degrees
  • Scanning angle: 52 degrees
  • Image depth (displayed image): 11.9 mm
  • Image depth (displayed image): 48 mm
  • Focal depth: 12.5 mm
  • Focal depth: 25 mm
  • Image width at focal zone: 15-18 mm

  • Image width at focal zone: 19-36 mm
  • Focal range: 10.5-14.5 mm
  • Focal range: 15-35 mm
 

A-Scan Modes

Axial Length Biometry A-Scan

Standardized Diagnostic A-Scan

  • Immersion or contact method
  • Movie sequence adjustable up to 5 seconds
  • Solid focused probe with internal fixation light

  • 50 frames-per-second image acquisition rate

  • Probe frequency: 10 MHz
  • Two caliper measurements displayed in mm with variable velocities

  • Image depth: 40 mm
  • Tissue sensitivity value stored in memory with reset function

  • Points on x-axis: 2048
  • Probe frequency: 8 MHz parallel beam

  • 8 bit resolution

  • Measurement accuracy: 50 microns inherent, 100 microns clinical
  • Steps of resolution: 256
 
  • Measurement accuracy: 50 microns inherent, 100 microns clinical
 
  • Automatic or manual scan acquisition
 
  • Built-in pattern recognition with automatic scleral echo detection

 
  • Statistics: average and standard deviation
 
  • Movie sequence adjustable up to 5 seconds
 
  • 50 frames-per-second image acquisition rate
 
  • IOL power calculations and analysis:

 
  • Holladay-I
 
  • SRK-T
 
  • Haigis
 
  • Hoffer-Q

 
Network and Connectivity
  • USB connectivity to off-the shelf Windows® Notebook computer (*)
  • New, easy-to-use GUI (graphical interface)
  • Multilingual user interface
  • Ellex remote access
Notebook Requirements
  • Processor: Quad Core™ Intel® i7
  • RAM: 8GB
  • Operating System: Windows® 10 Professional (64bit)
  • Display: 15.6 full HD (1920 x 1080)
  • Graphics: 2GB or higher video memory
  • Hard Drive: 512GB or larger
Data Management
  • Data archiving and image export capability
  • Customized report capability
  • DICOM connectivity
  • Verification of multiple concurrent DICOM connections to other Application Entities (AEs)
  • Query / retrieval of modality work list (patient data from Electronic Medical Records - EMR)
  • Storage of DICOM objects to EMR / Picture Archiving and Communication Systems (PACS)
Hardware Features
  • Slim line console with removable probe holders for easy cleaning
  • Footswitch control (scan start; scan stop)
Electrical Requirements
  • Power supply : 100-240 VAC auto-ranging
  • Frequency : 50/60 Hz
  • Input power : 50 VA
  • System Size (excl. computer) : 19 x 11 x 6 inches (47.5 x 27 x 15 cm)
  • Weight (excl. computer) : 11 lbs. (5 kg)

Specifications are subject to change without notice.

Accessories

Probes

A-Scan Probes

8 MHz probe for standardized diagnostic

a-scan_02

A-Scan Probes

10 MHz focus probe with internal fixation light for biometric A-Scan

a-scan_01

B-Scan Probes

10 MHz sealed B-Scan probe

b-scan_03

B-Scan Probes

40 MHz UBM Wide-Field Anterior Segment

B Scan Probes

Flanged Shells for B-Scan

(Optional Accessories)

Scleral Shells for B-Scan

(Optional Accessories)

Available in:

20mm Adult size

18mm Pediatric size

Ellex 40Mhz Probe 35

ClearScan® Bag for 40MHz B-Scan

(Optional Accessory)

ClearScan® is an innovative single-use ultrasound probe cover. Consisting of an extremely thin film that is acoustically invisible, ClearScan® provides distortion free ultrasound imaging with the added benefit of patient comfort. In addition, the ClearScan® conical shape enables the
safe and effective examination of all eye quadrants without causing corneal abrasion.

*ClearScan is a registered trademark of ESI, Inc.

ClearScan Bag for 40MHz B-Scan (Optional Accessory)